Doğrusal ivmelendirici cihazlarının NCRP?151 raporuna göre zırhlanması
No Thumbnail Available
Files
Date
2012
Authors
Journal Title
Journal ISSN
Volume Title
Publisher
Fen Bilimleri Enstitüsü
Abstract
In this thesis, it has been given the information about materials used and shielding calculations for linear accelerator devices (linac) according to ALARA (As Low As Reasonably Achievable) principle.Shielding calculations have been made for Electa Synergy device. The expected workload for 18 MV x-rays is 5 days per week, 8 hours per day, 20 patients per 8 hours and the value of absorbed dose at isocenter for each patient is 3 Gy. It is anticipated that an additional 30 patients per day will be treated with 6 MV x-rays to the same absorbed dose. P (shielding design goal) is taken (0,02 mSv/week) 20x10^(-6) Sv/week for uncontrolled areas and (0,1 mSv / week) 0,1x10^(-3) Sv / week for controlled areas.As a result of the calculations made, it has seen that the device is appropriately shielded and the measurements taken outside of the room are at the acceptable levels.
Description
Keywords
Doğrusal ivme, NCRP?151 raporu